When a curing light emits an intensity less than 300 mW/cm^2, checkUP provides a low output notification, to ensure the clinician is aware of the performance of their light in comparison to the expectation set by the manufacturer. According to ISO 10650:2018 light output shall vary by no more than +/- 20% of the manufacturer's stated output.
CheckUP does not indicate curing times for lights emitting less than 300 mW/cm^2 as a minimum. A light curing unit should also meet the minimum irradiance required by a dental material.
Omidi, Baharan-Ranjbar, et al. "Intensity output and effectiveness of light curing units in dental offices." Journal of clinical and experimental dentistry 10.6 (2018): e555.
Fan, P. L., et al. "Curing-light intensity and depth of cure of resin-based composites tested according to international standards." The Journal of the American Dental Association133.4 (2002): 429-434.
International Organization for Standardization. (2018). Dentistry — Powered polymerization activators (ISO Standard No. 10650:2018(E)). Retrieved from https://www.iso.org/standard/73302.html